AquaDMS - Water Disinfection Measuring System


Water disinfection monitor with touch screen colour display

The measuring system includes everything needed for disinfectant measurement: instrument, sensor, assembly and wiring – pre-mounted on a measuring panel. The system is used for measuring one of the following parameters: free Chlorine, Chlorine dioxide, Ozone or Hydrogen Peroxide. Measuring parameter and measuring range can be chosen in the menu. The integrated armature ensures a constant flow of approx. 30 l/h. For measuring free Chlorine, Chlorine Dioxide or Ozone a double gold sensor is used, for measuring in brine a special graphite-platinum sensor is available. For the measurement of Hydrogen Peroxide, a double platinum sensor is used. Temperature and flow is guarded by a further sensor, which additionally can be used for grounding. The control unit with touch screen operation includes an input for the measurement and temperature, one digital input, one analogue output and alarm relay, as well as an automatic sensor cleaning. Optionally, a system with integrated pH-measuring electrode is available which compensates pH fluctuations. The complete system is pressure resistant up to 6 bars at 20°C.

Application

Beer & Beverages

Features & Benefits

  • Complete, pre-mounted measuring panel.
  • Stabilized water flow
  • Fast results, measurement is directly made in the medium
  • Automated, patented sensor cleaning
  • Configuration without or with pH compensation
  • No zero drift
  • Control unit with touch screen
  • SD card for loading and saving setup, saving of diagnosis data and software updates
Specifications

  • Measuring principle: Potentiostatic measurement
  • Measuring span: Free Chlorine: 0 .. 20 mg/l
  • Chlorine Dioxide: 0 .. 20 mg/l, upon request 0 .. 30 mg/l
  • Ozone: 0 .. 20 mg/l
  • Hydrogen Peroxide: 0 .. 30 mg/l
  • Measuring range: Freely programmable except for H2O2, Standard 0 .. 5 mg/l
  • Measurment precision: +/- 2 % full scale
  • Resolution: 0.01 mg/l
  • Sample temperature: 0 °C .. +50 °C
  • Maximum pressure: 6 bar @ 20 °C
  • Conductivity of sample: minimum 50 μS/cm
  • pH of sample: 6 .. 9 (for free Chlorine 6 .. 8)
  • Ambient temperature: 0 °C .. +50 °C
  • Control unit:
    Outputs:  1 – 5 . 0/4 .. 20 mA
    1 Relay 250 VAC, 4 A, (NO/NC)
    Input: 1 . digital (NO/NC)
    Digital interface: microSD-card